MOSCOW BAUMANN STATE TECHNICAL UNIVERSITY
DEMO VERSION!!!

BACK

AIM OF THE WORK

SHORT THEORY

   Plasma

   Radiation by the Thermodynamic Equilibrium

   Radiation in the Conditions of the Local Thermodynamic Equilibrium

   Method of Relative Intensities

   Advisable literature

REFERENCE INFORMATION

   Spectral lines

   Argon (Ar I)

   Argon (Ar II)

   Crom (Cr I)

   Iron (Fe I)

   Hydrogen (H I)

   Mercury (Hg I)

   Calculation formulas

EQUIPMENT

   Spectrometer

   Optical behch design

   Optical resolution

   CCD-detector

   Fibre optic cable

   Collimating lens

   AVALight lamp

   SPECTROMETER CONTROL PARAMETERS

EXPERIMENT

   Experiment scheme

   Àcquaintance with control interfaces

   Order of carrying out the work

   Practical part
 
SPECTROMETER CONTROL PARAMETERS
For spectometer control are used next parameter groups.

  • Chanels and Smoothing - this options allows to switch on/off every optical chanel (Master, Slave1, Slave2, Slave3) and set smothing parameter in range from 0 to 7 separatedly for each chanel.

  • Scan parametrs:
    • integration time- this option changes the CCD readout frequency and therefore the integration time of the CCD detector (from 2 to 60000 msec);
    • average - number of scans to avarage;
    • saturation detection level .
  • Correct for Dynamic Dark - includes 2 parameters:
    • switch on/off correction procedure;
    • forget percentage - percentage of the new dark value pixels that has to be used (from 0% to 100%).