MOSCOW BAUMANN STATE TECHNICAL UNIVERSITY
DEMO VERSION!!!

INTERNET LABORATORY "PLASMA SPECTROMETRY AND PLASMA NANOTECHNOLOGIES"

   Log in

   Registration

   Password reset

   Contact informaton

EQUIPMENT

   Spectrometer

   Optical behch design

   Optical resolution

   CCD-detector

   Microprocessor board

   Fibre optic cable

   Collimating lens

   Objects of research

   AVALight lamp

   Halogen and deuterium lamps

SPECTROMETER CONTROL PARAMETERS

   Optical channels

   Smoothing

   Integration time

   Average

   Correct for dynamic dark

   Saturation detection levels

TRAININGS

   Technics and practice of spectrofotometry

   The qualitative analysis of radiation spectrum

   Definition of the Electronic Temperature in Plasma
 
Optical behch design
The fiber optic spectrometer AvaSpec 2048 are 2 optical benches with 75 mm focal length, developed in a symmetrical Czerny-Turner design.

Light enters the optical bench through a standard SMA905 connector (1) and is collimated by a spherical mirror (2). A plane grating (3) diffracts the collimated light; a second spherical mirror (4) focuses the resulting diffracted light. An image of the spectrum is projected onto a 1-dimensional linear detector array (5).

The optical bench has a number of components installed inside, allowing a wide variety of different configurations, depending on the intended application. The choice of these components such as the diffraction grating, entrance slit, order sorting filter, and detector coating have a strong influence on system specifications.

AvaBench-75 technical data
Focal length 75 mm
Numerical aperture 0.07
Stray light < 0.1 %
Detector SONY 2048
Dimensions, weight 120x91 x 21 mm, 255 gr.